Facilities

2016.09.30 14:51

Glove box (Hermetically Sealed System), Korea kiyon

Atachment
첨부파일 '1'
?

단축키

Prev이전 문서

Next다음 문서

+ - Up Down Comment Print
?

단축키

Prev이전 문서

Next다음 문서

+ - Up Down Comment Print
glove box.jpg

Specifications

1.  Hermetically Sealed Module (One Workstation) : 1 set
     Single sided and Single(2 port) length: 1,200(L) X 890(H) X 760mm(D)
     Modular structure with elastomer gasket front panel.
     Material : 304 Stainless Steel, Main body : 3 mm thick.
     Side end panel : 4mm thick, 3 micron filter : 2 ea
     Fluorescent Skylight, 9" Glove Port (M/C) : 2 ea., Butyl Rubber Glove (8" dia.) : 1 pair
     Window :Scratch & High impact resistant, Optically clear polycarbonate
     Electrical Outlet : Two duplex (110V, 220V), Glove Port cover : Protect atmosphere during 
glove replacement

     Service port feedthru : On request
     Leakage Certification : No detectable leak at ±8" W.C with helium spectrometer 
calibrated

2.  Antechamber : 1 set
     This chamber will be installed right and left side of glove box.
     Size : 380 dia. x 500 mm length, High vacuum design, O-ring sealed.
     Vacuum capacity : 1 x 10-3 torr, Doors, vertical action.
     Vacuum gauge : Bourdon tube dial type.
     Connection : 3/8" backfill ball valve.
     Leakage Certification : No detectable leaks at full vacuum with helium mass spectrometer 
calibrated.

3.  Mini Antechamber : 1 set
     This chamber will be installed right and left side of glove box.
     150mm (dia.) x 290mm (Length), Doors : Single center clamped, O-ring.
     Gauge : Broudon tube dial type, Vacuum capacity : 1 x 10-3 torr.

4.  Automatic Pressure Controller (Color Touch Panel) : 1 set
     ─5" to 5" W.C., Limits easily adjustable.
     Read pressure and set limits. Manual foot switch operate whether automatic control is on 
or off.

     Two solenoid valves are operated. Gas pressure : 10 psi to 30 psi, approximately 25 psi suggested

     1 set of Vacuum Pump : 200L/min...Intake Type : NW25. Ultimate pressure : 5 x 10-4 torr.

    Two Stage, Power 220V, 1 Phase.

5.  Dri Trap (Purification System) : 1 set

6.  Support Stand

7.  Moisture Analyzer : 1 set
     Model: DMT152
     Type: Thin-Film capacitive polymer sensor

8.  Oxygen Analyzer : 1 set
     Model: LD-450-C(Converter), SD40NW(Sensor Unit)

9.  Solvent Trap : 1 set
     Chemical Filter.

 

      -  Link : http://koreakiyon.co.kr/

 

 

Usage / Application

-  Store reactive materials

-  Do experiment under inert atmosphere 



  1. Transmission Electron Microscope, FEI (Tecnai G2-20)

  2. Powder X-ray Diffractometer, Rigaku (Ultima III), ...

  3. Electrochemical Analyzer, CH Instruments (750E)

  4. Glove box (Hermetically Sealed System), Korea kiyon

  5. Tube Furnace, DAIHAN Scientific (FT-830)

Board Pagination ‹ Prev 1 Next ›
/ 1

Designed by sketchbooks.co.kr / sketchbook5 board skin

나눔글꼴 설치 안내


이 PC에는 나눔글꼴이 설치되어 있지 않습니다.

이 사이트를 나눔글꼴로 보기 위해서는
나눔글꼴을 설치해야 합니다.

설치 취소

Sketchbook5, 스케치북5

Sketchbook5, 스케치북5

Sketchbook5, 스케치북5

Sketchbook5, 스케치북5